43
Setting
Filter
What is "Filter"?
When load value is unstable, set a low-path filter or the number of moving average to
stabilize the load value.
A low-path filter can remove instantaneous changes such as external noises and
stabilize load values.
The function of the number of moving average calculates the average of loads for the set
number to stabilize the load value.
How to set
1. Press MENU key.
2. Use
⋀
⋁
key to select
Condition Setting
.
3. Press SET key.
4. Use
⋀
⋁
key to select
Filter
.
5. Press SET key.
6. Use
<
>
key to select the appropriate low-path filter. (default: 100)
7. Press SET key and confirm the cursor has turned green.
8. Press SET key again.
9. Use
<
>
key to select the appropriate number of moving average. (default: OFF)
10. Press SET key and confirm the cursor has turned green.
11. Press SET key again to complete this operation.