guidance 

Electrostatic discharge 

(ESD) 

IEC 61000-4-2 

±6 kV contact 

±8 kV air 

±6 kV contact 

±8 kV air 

Floors should be wood, concrete or 

ceramic tile. If floor are covered with 

synthetic material, the relative 

humidity should be at least 30%. If 

ESD interfere with the operation of 

equipment ,counter measurements 

such as wrist strap, grounding shall 

be considered. 

Electrical fast 

transient/burst 

IEC 61000-4-4 

±2 kV for power supply lines 

±1 kV for input/output    lines 

 

   

Not applicable 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

Surge 

IEC 61000-4-5 

±1 kV differential mode. ±2 kV 

common mode 

 

 

 

Not applicable 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

Voltage dips, short 

interruptions and voltage 

variations on power 

supply input lines 

IEC 61000-4-11 

<5% UT   

(>95% dip in UT) 

for 0.5 cycle 

 

40% UT 

(60% dip in UT) 

for 5 cycles 

 

70% UT 

(30% dip in UT) 

for 25 cycles 

 

<5% UT   

(>95% dip in UT) 

for 5 sec 

Not applicable 

Mains power quality should be that of 

a typical commercial or hospital 

environment. If the user of the 

TL-100Drequires continued 

operation during power mains 

interruptions, it is recommended that 

the TL-100Dbe powered from an 

uninterruptible power supply or a 

battery. 

 

Power frequency (50Hz) 

magnetic field IEC 

61000-4-8 

3A/m 

3A/m 

Power frequency magnetic fields 

should be at levels characteristic of a 

typical location in a typical 

commercial or hospital environment.

NOTE    UT is the a.c. mains voltage prior to application of the test level. 

 

Guidance and manufacture’s declaration – electromagnetic immunity

The  

[EQUIPMENT or SYSTEM] i

s

 intended for use in the electromagnetic environment specified below. The customer 

or the user of   

[EQUIPMENT or SYSTEM] 

should assure that it is used in such an environment.

 

Immunity test 

IEC 60601 test 

level 

Complian

ce level 

Electromagnetic environment - 

guidance 

Summary of Contents for FT-F11

Page 1: ...1 Instruction for Use lnfrared Thermometer Contact Type FT F11 BT FT F21 BT FT F12 BT FT F22 BT FT F11 FT F21 FT F12 FT F22 UO8080 UO8080E...

Page 2: ...Replacement 7 Operation Precautions 8 Precaution for Measuring Temperature 9 Product Specification 10 Common question concerning forehead temperature 2 Product Overview This infrared Thermometer measu...

Page 3: ...Battery Last Memorized Temperature Display 5 Measuring 1 Ear Measuring Mode Figure 5 1 Suitable for FT F11 FT F11 BT FT F21 FT F21 BT FT F12 FT F12 BT FT F22 FT F22 BT UO8080 UO8080E Turn on the powe...

Page 4: ...table for FT F21 FT F22 FT F21 BT FT F22 BT UO8080E Turn on the power button key of the IR Thermometer and then all displayed segments appear briefly as like on Figure 5 2A the displayed screen appear...

Page 5: ...5 seconds until the displayed screen appears or then press the Scan button to switch to ear or forehead measuring mode 5 Memory recall 12 memories recall Press SCAN button key to display M on the disp...

Page 6: ...lcohol IPA to clean up before or after use Avoid multiple using by different peoples for cross infection Note 1 Any solution dilution and method other than mentioned above might cause reliability prob...

Page 7: ...attery type and battery life AAA 2pcs Suitable for FT F11 FT F21 FT F11 BT FT F21 BT UO8080 UO8080E 1 PC 2032 button battery Suitable for FT F12 FT F22 FT F12 BT FT F22 BT The typical service life of...

Page 8: ...e measure voice or other languages voice 2 When the measurement is finished the IR thermometer shall report the measured temperature data The Blue Tooth function model is FT F11 BT FT F12 BT FT F21 BT...

Page 9: ...he forehead area for temperature given off by the arterial blood supply under the skin without worrying about correct positioning The gentle scan is comfortable and not invasive Measurement method Nor...

Page 10: ...nt Clinical accuracy characteristics and procedures are available from the manufacturer on request STATEMENTS AND DECLARATIONS 1 MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC an...

Page 11: ...the receiver is connected Consult the dealer or an experienced radio TV technician for help 5 Guidance and manufacturer s delclaration Guidance and manufacture s declaration electromagnetic emission T...

Page 12: ...5 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hosp...

Page 13: ...EC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affect...

Page 14: ...738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23 333 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estima...

Page 15: ...ng instructions IP22 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Management Zone Changping Town Dongguan...

Reviews: