background image

Traceable

®

 is a registered trademark of Control Company   

©

2008  Control Company.  92-4134-00 Rev. 1 052008

 

 

 

 
SPECIFICATIONS 
Range:

 

–58 to 1999 °F (–50 to 1300 °C) 

Resolution:

 

0.1° or 1° (selectable) 

Accuracy:

 

±0.3% + 1°C 

Sampling Rate:

   2.5 times per second 

Display:

 

3½ digit LCD, ⅝ inches high (1.5cm) 

Fail-safe:

 

Low battery indicator 

Power:

 

9 volt alkaline battery 

Accessories:

 

Type-K thermocouple, rubberized case and a flip-out stand 

FRONT PANEL QUICK REFERENCE 
ON/OFF:

  Turns power on and off 

MAX:

 

Records and updates maximum values 

HOLD:

 

“Freezes” the display 

0.1°:

 

Selects resolution to 0.1° 

1°:

 

Selects resolution to 1° 

°C/°F:

 

Selects the temperature to be displayed in either Fahrenheit or Celsius  

OPERATION 

1. 

Insert the probe plug into the receptacle located on the front of the thermometer. 
The plug and receptacle are keyed with a small and a large pin. Make certain 
that the plug is inserted properly. 

2. 

Press the ON/OFF button to turn the unit on. 

3. 

Press the °C/°F button to display the temperature in Fahrenheit (°F will appear 
on the display) or Celsius (°C will appear on the display). 

4. 

Set the thermometer to desired resolution 0.1° or 1°. 

5. 

Place the probe tip in contact with the material to be measured and read the 
temperature on the display. 

MAX MODE 

In the MAX mode, the thermometer will display the highest temperature reading 
achieved. It does not display the current temperature if it falls below the MAX reading. 

CURRENT TEMPERATURE

 

DISPLAY READS

 

42° 

42° 

60° 

60° 

40° 

60° 

Example: In the MAX mode  

Press the MAX button (MAX will appear on the display). The thermometer will display 
the maximum temperature reading on the LCD display. Press the MAX button again to 
exit MAX recording mode. 

HOLD 

Press the HOLD button to “freeze” the display at the current reading (HOLD will 
appear in the display). Press the HOLD button again to return to the current reading. 

PROBES 

The probe supplied are ultra-fast response Type K thermocouple naked bead probes 
with 4’ leads and teflon insulation. Maximum operating temperature range for this 
thermocouple is 500°F (260°C). 

An accessory stainless steel triple purpose probe with handle is available. An 
accessory surface probe is also available (see accessories). 

Longer leads and extension may be used. The lead effect for 164 feet (50 meters) is 
generally less the 0.2°C with a limit of error of 2.2°C. Type-K wire and Type-K 
connectors must be used. 

RECORDER OUTPUT 

A voltage proportional to the display reading appears at the recorder OUTPUT socket 
located on the right side of the unit. Output is via a standard 3-pole 3.5 mm jack. 

 

 

 

Output specifications are as follows: 

1mV output per degree in the 0.1° display mode 
0.1 mV output per degree in the 1° display mode 

Output examples: 

Mode

 

Readings

 

Output

 

0.1° 

26.5° 

26.5 mV 

1° 

1500° 

150.0 mV 

OFFSET ADJUSTMENT (Fine tuning) 

The offset controls are set at the factory to allow for the variations found in standard 
thermocouples. By adjusting the offset controls, found on the side of the unit, you can 
optimize measurement accuracy for a particular thermocouple at a particular 
temperature. 

This unit has been calibrated Traceable

®

 to standards provided by the National 

Institute of Standards and Technology (NIST). While offset adjustments are possible, 
they will interfere with the factory calibration. Traceable

®

 recalibration must be made 

by the factory (see Recalibration). 

BENCH STAND/WALL MOUNT 

This unit is supplied with a bench stand that is on the back of the case. Locate the 
rectangular slots on the raised portion of the back of the unit. With your thumb and 
forefinger, lift the stand out. To close the stand, simply snap it shut.  

This unit may be wall mounted by using a screw. Set the screw into the wall at the 
location desired, the head of the screw will need to slip into the receptacle on the back 
of the unit, so do not set the screw flush to the wall. Once the screw has been properly 
set, hang the unit in place by sliding the receptacle on the back of the unit over the 
head of the screw. 

ACCESSORIES 
Control Cat. No. 4014— 

Stainless Steel Triple Purpose Probe, probe diameter

 1

/

8

" (0.32cm), probe length 6" 

(15.2cm), overall length 9" (22.9cm), cable length 50" (127cm). 

Control Cat. No. 4008— 

Surface Probe, flat disk on end has a diameter of 0.39" (1cm), overall length 9" 
(23cm), cable length 36" (91.4cm). 

Control Cat. No. 4028— 

Ultra-fast response, naked bead thermocouple. (identical to probe supplied with unit) 
cable length 48" (122cm).

 

Control Cat. No. 8039—

 

Low-Temperature Probe - Stainless-steel with handle; triple purpose (liquids, air/gas, 
and semi-solids), Dimensions: diameter 0.17 inch; stem length 12 inches; overall 
length 17 inches.  

Control Cat. No. 8613—

 

High-Temperature Probe - Ten-foot-long 0.19-inch diameter braided metal wire cable 
with smooth tip measures -73 to 982°C continuous or 1093°C short-term use.  
 

DISPLAY MESSAGES 
OL

 indicates that no probe is present indicates that the battery is low and needs 

replacement (see batter replacement). 

ALL OPERATIONAL DIFFICULTIES 

If this thermometer does not function properly for any reason, please replace the 
battery with a new high quality battery (see “Battery Replacement” section). Low 
battery power can occasionally cause any number of “apparent” operational difficulties. 
Replacing the battery with a new fresh battery will solve most difficulties. 

 

 

 

 