44
Testing FM Systems with FONIX 6500-CX Analyzers
Figure 2.1.2A
Leveling setup
Figure 2.1.2B
FM setup
2.1.3 Setup for Testing Environmental Microphones
If the client uses an environmental microphone with his FM system, it is important to test it
to make sure that its response matches that of the client’s hearing aid. This section describes
several different methods for setting up an environmental microphone for testing.
6 inches
FM transmitter
Probe mic and
FM receiver
Patient out of
sound field
6 inches
Reference mic