2.3 Configuration of transmitter ID and data transmission frequency
The measuring/transmission unit (12) is equipped with four configura-
tion switches for setting the transmitter ID and data transmission fre-
quency.
The switches illustrated are all OFF (actuator cams set to OFF).
2.3.1 Adjustment options of the transmitter detection
In the delivery condition all switches are set to OFF. It is recognised that,
in this case, bowl 1 and the transmission frequency of 1 second are the
set values. The bowl with ID 1 must then also be inserted in the indicated
position 1 of the mill. Similarly the EGTM system with ID 2 must be posi-
tioned in position 2 of the mill.
ID
Switch 1
Switch 2
Transmitter no. 1
OFF
OFF
Transmitter no. 2
ON
OFF
If other switch settings are set, the EGTM systems will not
be recognised. Ensure that the settings shown in the table
are used for your systems and that they are used in the cor‐
rect position in the mill.
EASY GTM Gas Pressure and Temperature Measuring System
- 42 -