Chapter 2—Daily Preparation for Treatment
2008K Hemodialysis Machine Operator’s Manual Rev P
45
4.
Verify that the concentrate type, displayed near the top of the screen, correctly matches
the prescribed concentrate type, and that the acid/bicarbonate or acetate concentrates
connected to the machine match the type selected. If an incorrect concentrate type is
displayed, the correct concentrate must be entered. To change the concentrate selection,
see “Setting an Acid/Bicarbonate Type” on page 66.
Figure 20 – The Dialysate Screen
Note
: If the machine is set up for use with Citrasate®, a ‘Citrate’ meter box will be
displayed in the dialysate constituent list (Functional board software version 3.36 or later).
5.
After the concentrate displayed is correct, verify that the Base Na+ and Bicarbonate are
as prescribed. Press the
CONFIRM
key, and then touch the
Home
screen-button.
6.
Insert the bicarbonate concentrate (blue) connector into a central bicarbonate supply or a
jug containing sufficient bicarbonate concentrate for an entire treatment. Again, be sure
the jug contains enough concentrate for the entire treatment
Preparing the Extracorporeal Blood Circuit
Use Figure 21 or Figure 22, depending on the configuration of your machine, as a guide for
connecting the bloodlines. The red lines on the machine are guides for arterial bloodline
(from patient to dialyzer). The blue lines on the machine are guides for the venous bloodline
(from the dialyzer to the patient). Be sure to use aseptic technique for all bloodline
connections.
Current acid
concentrate
selection