background image

1.

   Press ‘reinfusion’ and follow on screen instructions.

2.

   Disconnect the SafeLine from the post or pre-dilution connector 

8

 and attach to the recirculation adapter. 

3.

  Disconnect arterial line from patient.

4.

   Once connected press ‘OK’ (only press ‘NaCl’ if you have run 

out of acid/bicarb and need to use saline to wash back).

5.

   The blood pump will stop automatically when clear fluid is

recognised by the optical detector 

H

. Press ‘continue reinfusion’ 

if more fluid is  equired and manually stop the blood pump.

6.

   Disconnect the patient then press ‘remove lines’ button  

on screen.

7.

  Drain the dialyser.

8.

   Remove all blood lines ensuring Substituate port 

D

 is  

firmly closed

9.

   Replace concentrate suction tube and remove the bi

b

ag

®

  

when empty.

10.    

Close EBM doors 

2

 – machine automatically goes into hot 

disinfection, or press ‘heat disinfection’ on main screen.

0462-FMC-5008-TS-Brief-OpGuide.indd   9

09/09/2016   11:43

Summary of Contents for 5008

Page 1: ...The 5008 Therapy System Brief Operating Guide 62 FMC 5008 TS Brief OpGuide indd 1 09 09 2016 11 4 ...

Page 2: ...ituate port E Online Rinse port F Post dilution connector G Pre dilution connector H Blood air optical detector I Blood Temperature Monitor BTM Caution The colour coding of the dialyser couplings may vary from that shown above please check your local policy colour coding 62 FMC 5008 TS Brief OpGuide indd 2 09 09 2016 11 4 ...

Page 3: ...n of the clear plastic alpha clip B into the line guide then connect the arterial line to the SafeLine 4 7 Connect the rinse connector 5 to the venous line 8 On completion of T1 test insert the substituate connector into the ONLINE Substituate port D and the rinse connector on the venous line into the ONLINE Rinse port E 9 Close the doors and attach the dialyser couplings 6 as instructed 10 Press ...

Page 4: ...id or blood it must be replaced immediately Any fluid that is in the line must not be forced back using a syringe as this may damage the filter and may lead to contamination of the system 3 If at any time during treatment blood has reached the transducer protector the system must be checked for contamination after completion of the treatment check windows on white face of transducer protector A CO...

Page 5: ...t with the level detector 2 If there is a leakage of fluid f om the Safeline arterial patient line connection during preparation it is essential to clean and dry both the venous bubble catcher and the level detector Failure to do so may result in incorrect reading of blood levels in the venous bubble catcher at any time throughout the treatment 62 FMC 5008 TS Brief OpGuide indd 5 09 09 2016 11 4 ...

Page 6: ...ate port E Online Rinse port F Post dilution connector G Pre dilution connector H Blood air optical detector I Blood Temperature Monitor BTM Caution The colour coding of the dialyser couplings may vary from that shown above please check your local policy colour coding 62 FMC 5008 TS Brief OpGuide indd 6 09 09 2016 11 4 ...

Page 7: ...feLine sub tubing to the post or pre dilution connector F G 4 Ensure the Rinse port E is closed close the EBM doors open clamps and confirm blood pump start 5 Dialysis commences when blood is sensed by the optical detector H below venous bubble catcher message start dialysis treatment press OK 6 Increase blood flow to the p escribed speed 7 7 UF starts automatically if data is already set 8 Hepari...

Page 8: ...tuate port E Online Rinse port F Post dilution connector G Pre dilution connector H Blood air optical detector I Blood Temperature Monitor BTM Caution The colour coding of the dialyser couplings may vary from that shown above please check your local policy colour coding 62 FMC 5008 TS Brief OpGuide indd 8 09 09 2016 11 4 ...

Page 9: ...automatically when clear fluid is recognised by the optical detector H Press continue reinfusion if more fluid is equired and manually stop the blood pump 6 Disconnect the patient then press remove lines button on screen 7 Drain the dialyser 8 Remove all blood lines ensuring Substituate port D is firmly closed 9 Replace concentrate suction tube and remove the bibag when empty 10 Close EBM doors 2 ...

Page 10: ...mation in the manufacturer s operating manual which should be read before operating the machine Additional information and help regarding alarms is available by pressing the question mark help key in the alarm warning box on the screen 62 FMC 5008 TS Brief OpGuide indd 10 09 09 2016 11 4 ...

Page 11: ...ir detected below the venous bubble catcher Venous bloodline not properly inserted in optical detector or if during treatment the level in the venous bubble catcher needs to be raised Check level and raise if necessary using the up arrow in blood system screen 3 Level dropped in venous bubble catcher Visually check the blood level before choosing the option to raise the level If the level is corre...

Page 12: ... sensor Check all connections to identify the leak 5 DIASAFE plus leakage alarm There has been a leakage in the DIASAFE compartment of 50ml of fluid Open the compartment clean the drip tray and sensor Check the DIASAFEs have been inserted correctly Treatment may have to be discontinued Contact Technicians for assistance 62 FMC 5008 TS Brief OpGuide indd 12 09 09 2016 11 4 ...

Page 13: ...een instructions use operating manual if necessary 8 Air handling procedure Disconnect patient and re circulate blood lines following the on screen instructions 9 Venous pressure too high reduce pressure manually Open right hand door attach 20ml syringe to venous bubble catcher unclamp and release pressure alarm should clear automatically 62 FMC 5008 TS Brief OpGuide indd 13 09 09 2016 11 4 ...

Page 14: ...n arrow reduce level of fluid in the venous bubble catche Press re check on the screen If necessary change venous transducer protector 11 Arterial pressure measurement incorrect usually in prime mode Check arterial pressure dome is correctly inserted If alarm persists remove arterial line check cleanliness of arterial pressure unit Insert new arterial line if necessary 62 FMC 5008 TS Brief OpGuide...

Page 15: ...nserted in the line guide or treatment has been discontinued without using the re infusion programme Follow on screen instructions 13 Haemoconcentration warning Caused by a rise in TMP of 60mmhg during HDF or HF Follow on screen instructions and reduce substitution rate as advised 62 FMC 5008 TS Brief OpGuide indd 15 09 09 2016 11 4 ...

Page 16: ...Notes 62 FMC 5008 TS Brief OpGuide indd 16 09 09 2016 11 4 ...

Page 17: ...Notes 62 FMC 5008 TS Brief OpGuide indd 17 09 09 2016 11 4 ...

Page 18: ...Notes 62 FMC 5008 TS Brief OpGuide indd 18 09 09 2016 11 4 ...

Page 19: ...Notes 62 FMC 5008 TS Brief OpGuide indd 19 09 09 2016 11 4 ...

Page 20: ...Notes 62 FMC 5008 TS Brief OpGuide indd 20 09 09 2016 11 4 ...

Page 21: ...Notes 62 FMC 5008 TS Brief OpGuide indd 21 09 09 2016 11 4 ...

Page 22: ...uthwaite Sutton in Ashfield Notts NG17 2HU England Telephone 44 0 1623 445100 Fax 44 0 1623 550807 Head Office Else K öner Straße 1 61352 Bad Homburg v d H Germany www fmc ag com FMC UK 11 09 706 62 FMC 5008 TS Brief OpGuide indd 22 09 09 2016 11 4 ...

Reviews: