Venous bubble catcher and level detector.
1.
It is important to ensure that the level detector
is clean and dry and that the venous bubble
catcher is correctly positioned so that it makes
good contact with the level detector.
2.
If there is a leakage of fluid f om the Safeline/
arterial patient line connection during
preparation it is essential to clean and dry
both the venous bubble catcher and the
level detector. Failure to do so may result in
incorrect reading of blood levels in the venous
bubble catcher at any time throughout
the treatment.