4. Extracorporeal Blood Module (EBM) Sensor alarm
There is a leakage of 20ml of blood or priming fluid in
the extracorporeal blood module.
Open the doors and clean the drip tray/sensor.
Check all connections to identify the leak.
5. DIASAFE
®
plus
leakage alarm
There has been a leakage in the DIASAFE compartment of
50ml of fluid
Open the compartment, clean the drip tray and sensor.
Check the DIASAFEs have been inserted correctly.
Treatment may have to be discontinued. Contact Technicians
for assistance.