10. Venous pressure measurement incorrect
If during prime – remove venous transducer protector and reconnect.
Press ‘repeat check’.
Most common cause is a flooded venous transducer p otector.
Press blood system screen and using down arrow reduce level
of fluid in the venous bubble catche . Press ‘re-check’ on the screen.
If necessary change venous transducer protector.
11. Arterial pressure measurement incorrect
(usually in prime mode)
Check arterial pressure dome is correctly inserted. If alarm persists,
remove arterial line, check cleanliness of arterial pressure unit.
Insert new arterial line if necessary.