90PF
Section 3
Dual Channel Programmable
Remote Operation
Filter/Amplifier Card
11
1784 Chessie Lane, Ottawa, IL 61350 • Tel: 800/252-7074, 815/434-7800 • FAX: 815/434-8176
e-mail: [email protected]
•
Web Address: http://www.freqdev.com
POS=xy
Sets the post-gain of an active card
channel.
Set the post-gain where x can be any number
from 0 to 2, and y can be 0 to 9. Leading
zeroes may be dropped.
SAV=n
Saves the parameter settings of a
card/channel to a specific configuration
from a list of configurations. This set will
contain the active card/channel parameter
value settings.
Save current parameters into a configuration
where n is a value from 1 to 9.
OFFSET=sxyy
Sets the dc offset of an active card
channel.
Set the offset where s is the sign NOT used
in positive numbers, x is a number from 0 to
1, and y is a number from 0 to 9. Numbers
are left justified. Leading zeroes may be
dropped.
CFG=n
Initializes the parameters of a card/channel
to a specific configuration from a list of
configurations. This set will configure the
active card/channel.
Restore parameters from a configuration into
active card parameters where n is a value
from 1 to 9.
Table 2 – 90PF Command Set