90PF 

Section 2 

Dual Channel Programmable 

Software Installation 

Filter/Amplifier Card 

 

    

     

 

1784 Chessie Lane, Ottawa, IL 61350 • Tel: 800/252-7074, 815/434-7800 • FAX: 815/434-8176 

e-mail: [email protected] 

 Web Address: http://www.freqdev.com 

hear relays clicking for certain button changes in I/O).  When finished, use the Quit button to exit the 
application.  Using RS232 remote operation, after quitting a remote session, press the ENTER button on the 
90IP to return to local mode or turn the power button off and on, to reset the instrument. 

ƒ

 

For GPIB remote operation: Connect a GPIB cable from the instrument to the PC, and turn ON (I) the 90IP. 

ƒ

 

Initiate the 90IP Instrument Control from the desktop icon: Select GPIB then CLICK on the Scan button:  
GPIB0::1::INSTR should appear in the “Instruments” display window.  CLICK on items as they appear in the 
windows until the 90PF GUI appears.  Verify that control of the PF card is functional.  Use the Quit button to 
exit application. 

 

2.3 - Uninstall 

Executing setup toggles between install and uninstall.  To uninstall all the 90IP related software from the target PC, 
you may use the 

CDFC-01

 disk.  Uninstall is initiated from CD setup folders.  Each time this is done, the software 

toggles between install and uninstall.  However, even though the NI libraries auto setup, they do not uninstall 
automatically.  The NI libraries must be uninstalled separately using Add/Remove Programs in Control Panel.  The list 
is as follows, NI IVI engine 1.83, NI-MAX 2.2.0, NI-488.2 1.70, NI VISA 2.6 Run-time Engine, and NI-VISA Server 
1.0.0f6.  
 
 

Summary of Contents for 90PF

Page 1: ...90PF Dual Channel Programmable Filter Amplifier Card USER S MANUAL...

Page 2: ...1 2 90PF Operation Instructions 2 1 3 Front Panel Command Structure 3 Section 2 Software Installation 2 1 CDFC 01 Installation Disk 5 2 2 Remote Software Installation 6 2 3 Uninstall 7 Section 3 Remo...

Page 3: ...ctions 3 To make a change press the ENTER button and the addressed character will blink 4 Turning the SELECT knob will alter the value of the desired character or mode 5 Pressing the ENTER button agai...

Page 4: ...ecified filter module Filter type number of poles and filter transfer function are configured upon power up after insertion of the slot card Type P Transfer Function FIL LP Low Pass 8 B Butterworth HP...

Page 5: ...d filtered output signal will exit through channel two 2 output BNC For B REJECT operation input signal must enter through both channel 1 and channel 2 input BNC s Filtered output signal will exit thr...

Page 6: ...iles used in developing a custom 90PF slot card GUI 2 1 3 Setup Directory for Instrument Control Software 90IP Instrument Control Directory This directory installs the LAUNCH SLOT CARD CONTROL softwar...

Page 7: ...click on Setup exe to install the 90IPxxx Function Cards Driver software Verify that files reside in the selected target PC directory Select the 90PF User Interface folder Double click on Setup exe t...

Page 8: ...the desktop icon Select GPIB then CLICK on the Scan button GPIB0 1 INSTR should appear in the Instruments display window CLICK on items as they appear in the windows until the 90PF GUI appears Verify...

Page 9: ...here the RS232 interface card is installed in your PC 6 CLICK on SCAN this will display an instrument list CLICK or highlight the desired chassis this will display an instrument list of slot cards ins...

Page 10: ...Values returned by the 90IP are sent as a string of ASCII characters terminated by lf line feed or 0A x Table 1 shows examples of 90PF commands Command Function FREQ n Query the frequency setting of t...

Page 11: ...set setting A value from 100mV to 100 mV for example 05 mV 09 mV 10 mV etc BP Sets the channel configuration of active card channels to band pass filter Configure two channels as a band pass filter BR...

Page 12: ...configurations This set will contain the active card channel parameter value settings Save current parameters into a configuration where n is a value from 1 to 9 OFFSET sxyy Sets the dc offset of an a...

Page 13: ...vided into three functional groups Command Function and Configuration The Command Group allows the user to initiate certain process system commands The Function Group presents the filter type containe...

Page 14: ...Single Ended for single ended input operation or Differential for differential input mode 6 Output Configuration Control Use this control to select the output configuration for the channel on the slo...

Page 15: ...installation and have it work regardless of privileges As an administrator after the installation The account with administrator privileges that installed the software receives full control of the 90...

Page 16: ...t Control window Select the Users group and allow full control by checking on the Full Control box CLICK OK 18 Right click on the Measurement Automation shortcut and select Properties from the popup m...