90PF Slot Card 

 

Section 1 

Dual Channel Programmable 

Front Panel Operation 

Filter/Amplifier Card 

 

     

     

 

1784 Chessie Lane, Ottawa, IL 61350 • Tel: 800/252-7074, 815/434-7800 • FAX: 815/434-8176 

e-mail: [email protected] 

 Web Address: http://www.freqdev.com 

11. CHAN CFG

 – For slot cards that have a high-pass 

filter in channel 1 and a low-pass filter in channel 2, 
Channel Configuration (CHAN CFG) allows each 
channel to operate independently (DUAL) or to 
operate in a band-pass or band-reject mode. 

DUAL 

Utilize each channel independently 

*B-PASS 

Band Pass Operation 

*B-REJECT 

Band Reject (Notch) Operation 

*Band Pass (B-PASS) and Band Reject (B-REJECT) 
can be selected during configuration of channel 1 or 
channel 2.  The slot card must have a high pass filter 
in channel 1 and a low pass filter in channel 2.  For 

B-

PASS operation

, input signal must be connected to 

channel one (1) input BNC and filtered output signal 
will exit through channel two (2) output BNC.  For 

B-

REJECT operation

, input signal must enter through 

both channel (1) and channel (2) input BNC’s.  Filtered 
output signal will exit through channel two (2) output 
BNC. 

 
12. CONFIG

 – Up to 9 configurations can be saved on 

each 90PF slot card.  Slot card configuration, in use at 
time of power-OFF, will be saved to the next power 
ON cycle. 

CONFIG 

Select 

1 thru 9

 

 

S

 - Save

d

 

 

13. REMOTE 

– Displays Local reset screen when 

external computer has control of the instrument in 
remote operation.  On power-up, Instrument defaults 
to local mode. During a remote session, press ENTR 
to return to local mode. An external computer can only 
initiate remote mode. 

REMOTE

 

LCL

 - Local 

ENTR

 – RS232 or GPIB 

 

14. LED’s

 – Each 90PF contains two clipping sections 

per channel, one for the pre-gain output and one for 
the post-gain output.  These outputs are the inputs into 
the clipping circuitry.  The inputs pass through a 
comparator.  When peak voltage is greater than 
+10.5V or less than –10.5V, 

CLIPPING

 occurs and the 

red I/O LED’s will light. 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for 90PF

Page 1: ...90PF Dual Channel Programmable Filter Amplifier Card USER S MANUAL...

Page 2: ...1 2 90PF Operation Instructions 2 1 3 Front Panel Command Structure 3 Section 2 Software Installation 2 1 CDFC 01 Installation Disk 5 2 2 Remote Software Installation 6 2 3 Uninstall 7 Section 3 Remo...

Page 3: ...ctions 3 To make a change press the ENTER button and the addressed character will blink 4 Turning the SELECT knob will alter the value of the desired character or mode 5 Pressing the ENTER button agai...

Page 4: ...ecified filter module Filter type number of poles and filter transfer function are configured upon power up after insertion of the slot card Type P Transfer Function FIL LP Low Pass 8 B Butterworth HP...

Page 5: ...d filtered output signal will exit through channel two 2 output BNC For B REJECT operation input signal must enter through both channel 1 and channel 2 input BNC s Filtered output signal will exit thr...

Page 6: ...iles used in developing a custom 90PF slot card GUI 2 1 3 Setup Directory for Instrument Control Software 90IP Instrument Control Directory This directory installs the LAUNCH SLOT CARD CONTROL softwar...

Page 7: ...click on Setup exe to install the 90IPxxx Function Cards Driver software Verify that files reside in the selected target PC directory Select the 90PF User Interface folder Double click on Setup exe t...

Page 8: ...the desktop icon Select GPIB then CLICK on the Scan button GPIB0 1 INSTR should appear in the Instruments display window CLICK on items as they appear in the windows until the 90PF GUI appears Verify...

Page 9: ...here the RS232 interface card is installed in your PC 6 CLICK on SCAN this will display an instrument list CLICK or highlight the desired chassis this will display an instrument list of slot cards ins...

Page 10: ...Values returned by the 90IP are sent as a string of ASCII characters terminated by lf line feed or 0A x Table 1 shows examples of 90PF commands Command Function FREQ n Query the frequency setting of t...

Page 11: ...set setting A value from 100mV to 100 mV for example 05 mV 09 mV 10 mV etc BP Sets the channel configuration of active card channels to band pass filter Configure two channels as a band pass filter BR...

Page 12: ...configurations This set will contain the active card channel parameter value settings Save current parameters into a configuration where n is a value from 1 to 9 OFFSET sxyy Sets the dc offset of an a...

Page 13: ...vided into three functional groups Command Function and Configuration The Command Group allows the user to initiate certain process system commands The Function Group presents the filter type containe...

Page 14: ...Single Ended for single ended input operation or Differential for differential input mode 6 Output Configuration Control Use this control to select the output configuration for the channel on the slo...

Page 15: ...installation and have it work regardless of privileges As an administrator after the installation The account with administrator privileges that installed the software receives full control of the 90...

Page 16: ...t Control window Select the Users group and allow full control by checking on the Full Control box CLICK OK 18 Right click on the Measurement Automation shortcut and select Properties from the popup m...

Reviews: