WLA200 Rev140407
21
Full Tank Correction
This fine calibration procedure is used to fine tune the tank display after a non-
linear or linear calibration has been completed. The calibration process requires that
the tank be full at the start of the procedure and the display shows 75% or more.
The term ‘swipe’ means to move the magnet up to and then away from the
CAL.
sensor at the bottom of the display.
To exit the calibration mode swipe the magnet eight times. The top eight display
LEDs go on, the calibration program terminates, no program data is changed.
1. Fill the tank. (Do not fill up into the fill tower.)
2. Apply power to the display module.
Note:
Once the calibration program is activated there is a six second time
out. Do not wait more than six seconds to move from step 3 to step 4.
3. Swipe the magnet five times to enter the calibration mode.
Result: Top five display LEDs flash on and off.
4. After three seconds swipe the magnet five times to set the calibration into
memory.
Result: Top five display LEDs go on.
5. After six seconds the calibration process will terminate and all the LEDs will
go on to show that the tank is full.
Summary of Contents for TankVision WLA200
Page 30: ...WLA200 Rev140407 30 NOTES...
Page 31: ...WLA200 Rev140407 31 NOTES...
Page 32: ...WLA200 Rev140407 32...