Fig. 55
Example of Communication error
Failure to connect to the FossConnect Network server could be due to a number of
reasons, e.g., internet unavailability, firewalls in local IT-systems, blocked IP-
address ranges and port number in local Networks, etc. Contact the IT support team
for the installation site in question (or a person with similar responsibilities and
knowledge about the local IT system).
Also, check the time and date settings on the local PC. If the time differs too much
between the local PC and the FossConnect server, it will not allow you to connect.
LinkDiagnostic
is a special FossConnect server only used for trouble shooting and
backup of QA samples. It is pre-filled and it is only necessary to tick the
Active
box (which can be found on
Network
properties dialogue when clicking the
Edit
button).
On the
Import
tab you can import a pre-made FossManager configuration file
(*.mcf), with prediction models, parameter profiles, products, etc., as well as
licenses for both prediction models and software features.
Fig. 56
For more options use the Local Configurator, where you can create new or edit
existing configurations.
Summary of Contents for NIRS DS3
Page 1: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 12: ......
Page 14: ...Fig 2 Front view Fig 3 Rear view 1 1 2 3 4 ...
Page 52: ...2 Enter the 4 digit code default 1234 Fig 71 Fig 72 3 Click Yes to continue Fig 73 ...
Page 80: ......
Page 82: ......
Page 83: ...See 600968671 Parts Accessories and Consumables on the USB with documentation ...
Page 84: ......
Page 85: ......
Page 86: ...Contact your local FOSS representative for information ...