User Manual 6008 1286 / Rev. 8
4:29
Example:
Fig. 4:47
Running a Check Sample definition (ISIscan Nova)
The check sample definition will every time it is performed re-define the target for
the check sample but it will not change the limit-settings (these can only be changed
in FossManager).
1. Select the check sample product and tick the box on the right hand side so the
Product name changes to "Check Sample (Definition)"
Fig. 4:48
2. Place the check sample in the instrument using the small cup holder.
3. Close the lid.
4. Press Start to analyze the check sample. The check sample will now go through
a series of analysis. ISIscan Nova may ask you for a Sample number (depending
on settings in Operation profile. A Sample number is not required, but if you
have more than one check sample it could be wise to enter the Serial number of
the check sample in this field for future reference.
5. Result for the check sample definition is displayed after analysis and this target
is stored in ISIscan Nova as the target value for the check sample test.
Running a Check Sample test (ISIscan Nova)
1. Select the check sample product (box to the right should not be ticked).
Fig. 4:49
2. Place the check sample in the instrument using the small cup holder.
3. Close the lid.
Summary of Contents for NIRS DA1650
Page 20: ...NIRS DA1650 2 12 User Manual 6008 1286 Rev 8 ...
Page 26: ...NIRS DA1650 3 6 User Manual 6008 1286 Rev 8 Fig 3 7 9 Press Install to start installation ...
Page 30: ...NIRS DA1650 3 10 User Manual 6008 1286 Rev 8 ...
Page 44: ...NIRS DA1650 4 14 User Manual 6008 1286 Rev 8 Fig 4 25 Fig 4 26 Fig 4 27 ...
Page 64: ...NIRS DA1650 5 4 User Manual 6008 1286 Rev 8 ...
Page 68: ...NIRS DA1650 6 4 User Manual 6008 1286 Rev 8 ...