4:12
User Manual 6008 1286 / Rev. 8
Instrument Calibration
1. Place the EWC tool in the instrument with the help of the small cup holder and
close the lid.
2. Insert the EWC USB key in the USB port on the back panel.
3. Select
Instrument Calibration
and press
Calibrate
.
Fig. 4:20
4. Running instrument calibration requires you to enter the 4-digit password
"1234" (preventing un-intentional start of the calibration sequence).
Fig. 4:21
5. If the EWC key is not properly connected to the instrument, a message will
appear.
Fig. 4:22
6. Press
Continue
and wait for calibration to complete.
Summary of Contents for NIRS DA1650
Page 20: ...NIRS DA1650 2 12 User Manual 6008 1286 Rev 8 ...
Page 26: ...NIRS DA1650 3 6 User Manual 6008 1286 Rev 8 Fig 3 7 9 Press Install to start installation ...
Page 30: ...NIRS DA1650 3 10 User Manual 6008 1286 Rev 8 ...
Page 44: ...NIRS DA1650 4 14 User Manual 6008 1286 Rev 8 Fig 4 25 Fig 4 26 Fig 4 27 ...
Page 64: ...NIRS DA1650 5 4 User Manual 6008 1286 Rev 8 ...
Page 68: ...NIRS DA1650 6 4 User Manual 6008 1286 Rev 8 ...