4:20
User Manual 6008 1286 / Rev. 8
4.3.11 Configuration
The
Configuration
view contains settings which normally are done at time of
installation but they can also be edited afterwards.
1. Press
Configuration
.
Fig. 4:35
On the
FossManager Configuration
tab you enter the Address and Port number for
the FossManager Network server to be used for this particular instrument. This
information will be provided to you by the Network manager or the FOSS
representative at time of installation.
Test FossManager connection
will test and verify if the connection to the
FossConnect server can be established.
Should an incorrect address or port number be entered or the Internet connection does
not allow for the Instrument to contact the server an error message will appear.
Fig. 4:36
Failure to connect to the FossManager Network server could be caused by a number
of different reasons. E.g. Internet availability, firewalls in local IT-systems, blocked
IP-address ranges and port numbers in local Networks etc. Contact the IT support
team for the installation site in question (or a person with similar responsibilities and
knowledge about the sites IT system).
Also check the time, date and timezone settings in the embedded PC. If time differs
too much between the embedded PC and the FossConnect server it will not allow you
to connect.
Summary of Contents for NIRS DA1650
Page 20: ...NIRS DA1650 2 12 User Manual 6008 1286 Rev 8 ...
Page 26: ...NIRS DA1650 3 6 User Manual 6008 1286 Rev 8 Fig 3 7 9 Press Install to start installation ...
Page 30: ...NIRS DA1650 3 10 User Manual 6008 1286 Rev 8 ...
Page 44: ...NIRS DA1650 4 14 User Manual 6008 1286 Rev 8 Fig 4 25 Fig 4 26 Fig 4 27 ...
Page 64: ...NIRS DA1650 5 4 User Manual 6008 1286 Rev 8 ...
Page 68: ...NIRS DA1650 6 4 User Manual 6008 1286 Rev 8 ...