User Manual 6008 1286 / Rev. 8
1.3
Product Safety
Caution
The instrument is designed and tested for European (CE)
compliance. To ensure that this compliance is maintained,
connect only CE approved equipment. Connecting equipment
that is not CE approved may cause EMC incompatibility and
thereby affect the function of the instrument and other
equipment.
Caution
When nothing is connected the sealing caps should be mounted
to maintain dust protection.
Caution
The air circulation around the lamp cooling flange must not be
inhibited. Leave uncovered with sufficient space for air
circulation at all times.
Caution
Do not touch the lamp glass or reflector or let any rough surface
come into contact with the lamp glass. A microscopic scratch in
the glass might cause a lamp explosion later.
Caution
The lamp can be damaged by fingerprints and oily residues.
Cotton gloves must be used when replacing the lamp to avoid
any damages.