63
10-2-5. Inverting GPI Output Pulse Polarity
To invert the polarity of GPI output pulse, change
Polarity
from
Normal
to Invert.
This function is used in such cases such as when inverting a tally condition, or using output
devices that work in inverse logic.
When connecting to FA-AUX30 units, set Polarity to
Normal
.
10-2-6. Sending GPI Settings
After all GPI settings are finished, click
Apply
to send all settings to the FA-10DCCRU unit.
The following message indicates that GPI ports have successfully been set. Click
OK
to
close the dialog window.
The following message indicates that GPI settings have failed. Click
OK
to close the dialog
window. Verify the network connection and settings, then click
Apply
to reenter settings.
10-2-7. Pattern Load
The Pattern Load function allows you to assign GPIO functions in groups by loading patterns.
<Pattern Load Procedure>
(1) Click to select a tab from
GPI Port 1 - 10(FA-AUX30 Left Block), GPI Port 11 -
20(FA-AUX30 Center Block)
and
Port 21 - 30(FA-AUX30 Right Block)
.
(2) Click on the Pattern box to select a pattern from the dropdown menu.
(3) Click
Load
to load the pattern. Ten GPI input/output ports are quickly set. Loaded settings
can be changed in the same manner as that of assignments. (See section 10-2-3.
“GPI
Inp
ut Settings” and section 10-2-4. “GPI Output Settings.”)
(4) After all GPI settings are finished, click
Apply
to send the settings to the FA-10DCCRU
unit. If
a “Successful settings” message appears, GPI settings are complete.
See section 10-2-8.
“GPI Pattern List” for details on GPI patterns.