Appendix B. Troubleshooting Tips
B-1. Introduction
In the event you experience a problem with the FONA CDRelite
Sensor or Interface Module, refer to the table of troubleshooting tips
found on this page. If the problem persists, please contact your local
distributor.
B-2. Troubleshooting Table
Item
Description
Corrective Action
1
Both green LEDs are off when CDR
DICOM (or other imaging software) is not
running.
Cable short or over-current condition. Replace Sensor cable.
2
The green LED on the Sensor side is blinking,
whether CDR DICOM (or other imaging
software) is running or not.
Low current condition. Replace Sensor cable.
3
Both green LEDs are off, whether CDR
DICOM (or other imaging software) is
running or not.
Over current condition. Replace Sensor cable.