FONA CDRelite System User Guide
B1057003 Rev. 3
43
A-5. EMC Tables
The following tables provide FONA CDRelite USB Interface
Module compliance information to electromagnetic compatibility
(EMC) and electromagnetic immunity (EMI) standards. To ensure
conformance, the customer or user must use the FONA CDRelite
USB Interface Module in environments that are consistent with
these standards.
The USB cable required with the FONA CDRelite USB Interface
Module must also comply with the same standards. The operation of
the FONA CDRelite USB Interface Module has been independently
tested using USB cables identified in Section 2.1.
Compliance to
EMC and EMI standards cannot be guaranteed by the use of
alternate cables.
Table 8. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
PLEASE NOTE: The FONA CDRelite USB Interface Module is intended for use in the electromagnetic environment
specified below. The customer or user of the FONA CDRelite USB Interface Module must ensure that it is used in such
an environment.
Emissions Test
Compliance
Guidance
RF emissions
CISPR 11
Group 1
The FONA CDRelite USB Interface Module uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The FONA CDRelite USB Interface Module is suitable for use
in all establishments including domestic establishments and
those directly connected to the public low-voltage supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 9. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
PLEASE NOTE: The FONA CDRelite USB Interface Module is intended for use in the electromagnetic environment
specified below. The customer or user of the FONA CDRelite USB Interface Module must ensure that it is used in such
an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 610004-4
2 kV for power supply
lines
1 kV for input/output
lines
2 kV for power
supply lines
1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2kV common mode
1 kV differential
mode
2kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.