H300P Temperature Humidity Chamber
Page 3
CAUTION
: This chamber is designed for operation in a conditioned laboratory environment. Operation
above 30°C (85°F) or below 16°C (60°F) ambient room temperature is NOT recommended.
Operation Safety Notices
CAUTION
: The Watlow “PM PLUS 6 User’s Manual” is a general manual and is written for a wide variety
of applications and configurations. Not all features or functions are applicable.
CAUTION
: The PM PLUS 6 Controllers have been properly configured at our factory to match the
chamber’s system requirements and to perform optimally over a wide range of operating
conditions. Improper modifications to these setup values can result in erratic performance
and unreliable operation. Do not attempt to modify the setup values, unless you thoroughly
understand what you are doing. If there is any doubt, please contact the Folyon
Technologies before proceeding.
CAUTION
: Avoid touching the sensor assembly inside the chamber.
WARNING
: Do NOT put items in the chamber that could burn or explode. This is NOT an explosion-proof
chamber.
WARNING
: Do NOT put items in the chamber which can emit corrosive vapors or substances.
WARNING
: This chamber is NOT a curing oven. There are NO provisions for venting fumes.
WARNING
: Always unplug the chamber from the outlet when filling the water reservoir with water. If
water is spilled inside the chamber, do not plug in! Allow sufficient time (at least 24 hours)
for the water to dry.
WARNING
:
The chamber is designed to operate in not condensing conditions only. It is up to the
operator to ensure that condensation does not occur inside the chamber. Condensation
inside the chamber can damage the sensors and the internal components of the chamber.
Operating the chamber when condensation is present inside the chamber will void the
warranty!
WARNING
:
Before the chamber is shipped using a common carrier such as UPS, FexEx, DHL and the
Postal Service, all water must be emptied from the water reservoir. The water can be
emptied by tilting the chamber and emptying all the water into a bucket, through the
filling tube. Make sure the unplug the chamber before filling or emptying water.