H300P Temperature Humidity Chamber
Page 12
Linearization Menu
For the relative humidity controller there are no changes to controller default parameters. For the
temperature controller, each H300P chamber is supplied with its own set of Linearization Menu
settings. These settings are included in a document on the flash drive that comes with each
chamber.
Process Value Menu
No changes to controller default parameters.
Control Loop Menu
Most of the parameters in the Control Loop Menu are specific to each H300P chamber. Each H300P
chamber is supplied with its own set Control Loop Menu settings. These settings are included in a
document on the flash drive that comes with each chamber.
Output Menu
Output 1
Parameter
Temperature
Humidity
Output Function
Heat
Heat
Time Base Type
Fixed Time Base Fixed Time Base
Fixed Time Base
1.0
3.0
Low Power Scale
0
0
High Power Scale
100
50
Output 2
Parameter
Temperature
Humidity
Output Function
Cool
Cool
Time Base Type
Fixed Time Base Fixed Time Base
Fixed Time Base
1.0
3.0
Low Power Scale
0
0
High Power Scale
100
100
Alarm Menu
No changes to controller default parameters.
Function Key Menu
No changes to controller default parameters.