H300P Temperature Humidity Chamber
Page 11
•
Press the Home key when done.
To turn the control back On follow these steps:
•
Press the Forward Arrow key once.
Control Mode 1
should be highlighted.
•
Press the Forward Arrow key once.
•
Press the
+
or
–
key until
Auto
is highlighted.
•
Press the Home key when done.
Setting Up a Profile (PM6R Models Only)
For instructions on how to set a profile, see the “PM Plus User’s Manual”, Chapter 6: Profiles. For easy
profile setup download the Watlow EZ-LINK™ App to your phone or tablet. The App communicates with the
controllers using Bluetooth®.
Setup Page Parameters
The setup page parameters below have been programmed into the temperature and humidity controllers at
the factory. These parameters are for reference only and should not be modified.
Improper modifications to these setup values can result in erratic performance and unreliable operation. Do
not attempt to modify the setup values, unless you thoroughly understand what you are doing.
These parameters were saved at the factory under User Settings 1. If these parameters are modified, they
can be restored to factory defaults. Please see the
Restoring Factory Default Parameters
manual on how to do this.
Analog Input Menu
Parameter
Temperature
Humidity
Sensor Type
RTD 100 Ohm
Milliamps
RTD Leads
3
Units
Relative Humidity
Scale Low
4.00
Scale High
20.00
Range Low
0.0
Range High
100.0
Process Error Enable
Off
Filter
0.5
0.5
Input Error Latching
Off
Off
Display Precision
Tenths
Tenths
Calibration Offset
0.0
0.0