background image

��

PAGE

40001
40002
40003
40004
40006
40008
40009
40010
40011
40012
40013
40014

uint16_t
uint16_t
uint16_t

float
float

uint16_t
uint16_t
uint16_t
uint16_t
uint16_t
uint16_t
uint16_t

read

RW
RW

read

RW
RW
RW
RW
RW
RW
RW
RW

Firmware version

Input signal

Output signal

Input value

Output value

Software output mode and output switch

Active current output value USER upper limit

Active current output value USER lower limit

Voltage output value USER upper limit

Voltage output value USER lower limit

Passive current output value USER upper limit

Passive current output value USER lower limit

Address

Type of Data

0X9C41
0X9C42
0X9C43
0X9C44
0X9C46

0X9c48

0X9C49
0X9C4A
0X9C4B
0X9C4C

0X9C4D
0X9C4E

Hexadecimal

Attributes

Illustrate

【  Table 2 】

Register upper 8 bits

Register lower 8 bits

01 is current; 02 is voltage; 03 is frequency; 04 is millivolt; 05 is resistance
The upper four bits are the type: 1-8 represent S, B, E, K, R, J, T, N in turn
Lower four-bit mode: 1 is mV, 2 is thermocouple, 3 is WR thermocouple

Enter

【  Table 3 】

100
101
102
103

0x64
0x65
0x66
0x67

read

write

read

write

Function code

Hexadecimal

uint16_t
uint16_t

float
float

Type of Data

Attributes

【 Table 1 】

【Example 】

slave

address

function

code

register

address

register address

low byte

Register number

high byte

CRC high

byte

CRC low

byte

01

65

9C

42

04

Register number

low byte

62

40

AF

Write input millivolt signal, J thermocouple type

Illustrate:

01: Slave address set by the device; 65: Write function code, see Table 1; 9C42: Read and write 

input signal address, see Table 2; 04: 01 is current, 02 is voltage, 03 is frequency, and 04 is 

millivolt , 05 is the resistor; 62: the upper four bits are the type: 1 is S, 2 is B, 3 is E, 4 is K, 5 is R, 6 

is J, 7 is T, 8 is N, the lower four bits mode: 1 is millivolt, 2 is thermocouple, 3 is WR 

thermocouple; 40AF: CRC code (note! Send in hex)

Summary of Contents for SG-003A

Page 1: ...Multi functional Signal Generator OPERATING MANUAL SG 003A...

Page 2: ......

Page 3: ...Signal 06 4 6 24V Loop Detection 07 4 Display Signal Description 2 1 Speci cations 01 2 3 Interface Terminal 02 2 2 Function Introduction 01 2 4 Power Charge 03 2 Product Description 5 1 Engineering...

Page 4: ...example the current signal of 4 20mA corresponds to 0 100 that is 4mA corresponds to 0 12mA corresponds to 50 and 20mA corresponds to 100 This range can be modi ed by parameters PAGE Plug the meter i...

Page 5: ...n a certain period of time Modbus Slave FNIRSI SG 003A can communicate with con guration software or PLC via RS485 set the instrument Firmware Upgrade USB emulated U disk for rmware upgrade When the I...

Page 6: ...version mode Button Function Operation Description RUN Light The power on on output is white light IN Input key Used to switch the input signal mode in the main interface When FN red light is on click...

Page 7: ...mode the button blue light at this time the up key single output the down key reset the left key end the output the right key pause continue ON OFF Con rmation button when the main interface Turn on...

Page 8: ...display of input output actual value and engineering quantity Output Open Circuit Detection When the output is turned on if the interface displays open circuit and ashes it means that the output sign...

Page 9: ...tion Operating Instructions For Speed Mode 1 The speed mode is used to calculate the speed of the motor or gear One revolution of the motor or gear produces 1 pulse one minute is 3000 revolutions so 1...

Page 10: ...upply and detect current 2 Loop 2 After activating the independent 24V power supply The IN port outputs 24V voltage between the 24V port and the COM port It does not a ect the detection type of the in...

Page 11: ...ude range of the output signal 3 Overrun allowed For example 1 5V is converted into 4 20ma If the user checks this option when the input is oV the output will follow to OmA If the user does not check...

Page 12: ...ys After starting the programming output mode the backlight of the up down left and right keys will turn blue The functions of the up down left and right buttons will also change that is the top is a...

Page 13: ...res Output Time Set output time 5 7 Communication USB connection Because the USB analog serial port is used for communica tion Note that charging will slow down after checking the USB connection Slave...

Page 14: ...ve position given Valve position feedback This input the port is closed to 24V as the negative pole of the input signal the input and output are performed at the same time PLC Process instrument etc A...

Page 15: ...e indicated by the technical indicator It may cause the instrument to work abnormally or even damage Error in input and output It is normal that there may be errors in input and output Because the inp...

Page 16: ...ignal Generator Brand Model FNIRSI SG 003A Manufacturer Shenzhen FNIRSI Technology Co Ltd Website www fnirsi cn Address Building C Weihuada Industrial Park Dalang Street Longhua District Shenzhen Guan...

Page 17: ...ge 03 is frequency 04 is millivolt 05 is resistance The upper four bits are the type 1 8 represent S B E K R J T N in turn Lower four bit mode 1 is mV 2 is thermocouple 3 is WR thermocouple Enter Tabl...

Page 18: ......

Page 19: ......

Page 20: ...Download Manuals and Firmware Manufacturer Shenzhen FNIRSI Technology Co Ltd Website www fnirsi cn Address Building C Weihuada Industrial Park Dalang Street Longhua District Shenzhen Guangdong China...

Reviews: