【 4.4 】  XMT Passive Current Signal

【 Mode Switch 】

:When the main interface 【 FN 】 is white.Press 【 IN/OUT 】 to switch 

the display of actual value and engineering quantity of input/output

【 Output Range 】

: The limit range of passive current output can be set

【 4.5 】  Hz Pulse Signal

【 Mode 】

: The pulse signal has four modes: frequency, PWM duty cycle, speed frequency, 

pulse quantitative/counting. When the main interface 【 FN 】 is white, long press 【 OUT 】 

to switch mode

【 Range 】

: Select the range of frequency output: 0~99.999Hz, 0~999.99Hz, 

0~9999Hz, 0~200kHz

【 Method 】

: Select level/open circuit according to the presence or absence of pull-up

【 Output Peak Value】

: The level height of the output waveform, the highest is 24V

【 Frequency Mode Operation Instructions 】

:The frequency range can be switched 

by pressing 【 OUT 】 when the main interface 【 FN 】 is white

【 Operating Instructions In PWM Mode 】

: PWM displays two parameters, the 

frequency value (that is, the period value) and the duty cycle.The frequency value 

will be displayed in the original engineering quantity position (no engineering 

quantity display).Note that the maximum frequency does not exceed 9999Hz.The 

frequency value/duty cycle parameters can be set when outputting.When the main 

interface 【 FN 】 is white, press 【 OUT 】 to switch, there is a yellow cursor indication

【 Operating Instructions For Speed Mode 】

:

(1)

 The speed mode is used to calculate the speed of the motor or gear. One 

revolution of the motor or gear produces 1 pulse, one minute is 3000 revolutions, so 1 

second is 50 pulses

(2)【 Time Unit 】

: Frequency=RPS(per second)*number of pulses per revolution, 

RPM(per minute)=RPS*60, RPH(per hour)=RPS*3600.In the speed setting selection, 

note that the maximum frequency does not exceed 9999Hz

(3)【 Number Of Revolution Pulses 】

: The number of pulses per revolution of the 

motor or gear, which can be selected in the speed setting

(4)【 Tooth Missing 】

: It is mainly used to simulate the engine speed, select it in the 

speed setting

【 Quantitative/Counting Mode Operation Instructions 】

: The main purpose of 

pulse quantitative is to simulate the quantitative control sensor of flow.the frequency 

value will be displayed in the original quantity position.The cycle is displayed at the 

original engineering position, and the unit is milliseconds.When the main interface 

【 FN 】 is white, press 【 OUT 】 to switch between cycle and quantitative values, which 

are distinguished by cursors
The period value can be modified during quantitative output.In this mode, when 

the cursor disappears after the output is turned on, the backlight of the up, down, 

left, and right keys will turn blue.The function of the up, down, left, and right 

buttons will also change, that is, the up is a single pulse sent.Down is program stop 

and reset, left is no function, right is start/pause

��

PAGE

Summary of Contents for SG-003A

Page 1: ...Multi functional Signal Generator OPERATING MANUAL SG 003A...

Page 2: ......

Page 3: ...Signal 06 4 6 24V Loop Detection 07 4 Display Signal Description 2 1 Speci cations 01 2 3 Interface Terminal 02 2 2 Function Introduction 01 2 4 Power Charge 03 2 Product Description 5 1 Engineering...

Page 4: ...example the current signal of 4 20mA corresponds to 0 100 that is 4mA corresponds to 0 12mA corresponds to 50 and 20mA corresponds to 100 This range can be modi ed by parameters PAGE Plug the meter i...

Page 5: ...n a certain period of time Modbus Slave FNIRSI SG 003A can communicate with con guration software or PLC via RS485 set the instrument Firmware Upgrade USB emulated U disk for rmware upgrade When the I...

Page 6: ...version mode Button Function Operation Description RUN Light The power on on output is white light IN Input key Used to switch the input signal mode in the main interface When FN red light is on click...

Page 7: ...mode the button blue light at this time the up key single output the down key reset the left key end the output the right key pause continue ON OFF Con rmation button when the main interface Turn on...

Page 8: ...display of input output actual value and engineering quantity Output Open Circuit Detection When the output is turned on if the interface displays open circuit and ashes it means that the output sign...

Page 9: ...tion Operating Instructions For Speed Mode 1 The speed mode is used to calculate the speed of the motor or gear One revolution of the motor or gear produces 1 pulse one minute is 3000 revolutions so 1...

Page 10: ...upply and detect current 2 Loop 2 After activating the independent 24V power supply The IN port outputs 24V voltage between the 24V port and the COM port It does not a ect the detection type of the in...

Page 11: ...ude range of the output signal 3 Overrun allowed For example 1 5V is converted into 4 20ma If the user checks this option when the input is oV the output will follow to OmA If the user does not check...

Page 12: ...ys After starting the programming output mode the backlight of the up down left and right keys will turn blue The functions of the up down left and right buttons will also change that is the top is a...

Page 13: ...res Output Time Set output time 5 7 Communication USB connection Because the USB analog serial port is used for communica tion Note that charging will slow down after checking the USB connection Slave...

Page 14: ...ve position given Valve position feedback This input the port is closed to 24V as the negative pole of the input signal the input and output are performed at the same time PLC Process instrument etc A...

Page 15: ...e indicated by the technical indicator It may cause the instrument to work abnormally or even damage Error in input and output It is normal that there may be errors in input and output Because the inp...

Page 16: ...ignal Generator Brand Model FNIRSI SG 003A Manufacturer Shenzhen FNIRSI Technology Co Ltd Website www fnirsi cn Address Building C Weihuada Industrial Park Dalang Street Longhua District Shenzhen Guan...

Page 17: ...ge 03 is frequency 04 is millivolt 05 is resistance The upper four bits are the type 1 8 represent S B E K R J T N in turn Lower four bit mode 1 is mV 2 is thermocouple 3 is WR thermocouple Enter Tabl...

Page 18: ......

Page 19: ......

Page 20: ...Download Manuals and Firmware Manufacturer Shenzhen FNIRSI Technology Co Ltd Website www fnirsi cn Address Building C Weihuada Industrial Park Dalang Street Longhua District Shenzhen Guangdong China...

Reviews: