Operating Manual EMGZ621A
22
Gain 1 B
Use:
Identical
with
Gain B
but the value was determined by special fuction
Calib. gain 1 B
. The value stored here is used if the dig. input „Switch gain
B“ is activated (ref. to wiring diagram, fig. 6).
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Force of sensor
Use:
This parameter stores the nominal force of the sensors. It is printed to the
nameplate of the force sensors.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N, kN, cN]
Unit of sensor
Use:
This parameter stores the measuring unit of the sensors. It is printed to the
nameplate of the force sensors.
Range:
N, kN, cN
Default:
N
Sensitivity
Use:
This parameter stores the sensitivity of the force sensors, that means how
much signal per volt excitation the sensor will give when loaded with
nominal force. Standard for FMS force sensors is 1.8mV/V.
Range:
0.1
to
5.0
Default:
1.8
Increment:
0.1
Unit:
[mV/V]
Min. limit
Use:
The digital output and the LED „Min. Limit A+B“ will be activated if the
tension feedback passes under the threshold value stored here.
If the parameter contains a zero value, limit switch monitoring is inactive.
Range:
A force value can be input. The value consists of 4 digits. The position of
the decimal point depends on the parameter
Force of sensor.
Default:
0
Unit:
[N, kN, cN]
or
[lb, klb, clb]
Summary of Contents for EMGZ621A
Page 8: ...Operating Manual EMGZ621A 8 6 Dimensions fig 3 Dimensions E611002e...
Page 11: ...Operating Manual EMGZ621A 11 7 2 Wiring Diagram fig 6 Wiring diagram EMGZ621A E621A0011e...
Page 26: ...Operating Manual EMGZ621A 26 10 Serial Interface RS232 Optional...
Page 31: ...Operating Manual EMGZ621A 31 13 Interface CAN Bus Option 14 Interface DeviceNet Option...