Operating Manual EMGZ621A
19
9.2
List of the system parameters
Parameter
Unit
Min
Max
Default
Selected
Language
English, French, Italian, German
Measuring System
Metric, US standard
Metric
Lowpass display
[Hz]
0.1
10.0
1.0
Identifier [-]
0
255
0
Baud rate
2400, 4800, 9600, 19200
9600
9.3
List of the parameters EMGZ621A
Parameter
Unit
Min
Max
Default
Selected
Offset A
[Digit]
–8000
8000
0
Gain A
[-]
0.100
9.000
1.000
Gain 1 A
[-]
0.100
9.000
1.000
Offset B
[Digit]
–8000
8000
0
Gain B
[-]
0.100
9.000
1.000
Gain 1 B
[-]
0.100
9.000
1.000
Force of sensor
[N, kN, cN]
1
9999
1000
Unit of sensor
N, kN, cN
N
Sensitivity [mV/V]
0.1
3.0
1.8
Min. limit
1)
2)
0
Max. limit
1)
2)
0
|A–B| > limit
1)
2)
-
Lowpass output 1 (A–B)
[Hz]
0.1
200.0
10.0
Scale output 1 (A–B)
1)
2)
-
Config. output 1 (A–B)
0...20mA, 4...20mA
0...20mA
Lowpass output 2 (A+B)
[Hz]
0.1
200.0
10.0
Scale output 2 (A+B)
1)
2)
-
Config. output 2 (A+B)
0...20mA, 4...20mA
0...20mA
Lowpass output 3 (A)
[Hz]
0.1
200.0
10.0
Scale output 3 (A)
1)
2)
-
Config. output 3 (A)
0...20mA, 4...20mA
0...20mA
Lowpass output 4 (B)
[Hz]
0.1
200.0
10.0
Scale output 4 (B)
1)
2)
-
Config. output 4 (B)
0...20mA, 4...20mA
0...20mA
1)
[N, cN, kN] if measuring system = metric
[lb, clb, klb] if measuring system = US standard
2)
A force value can be input. The value consists of 4 digits. The
position of the decimal point depends on the parameter
Force of sensor
Summary of Contents for EMGZ621A
Page 8: ...Operating Manual EMGZ621A 8 6 Dimensions fig 3 Dimensions E611002e...
Page 11: ...Operating Manual EMGZ621A 11 7 2 Wiring Diagram fig 6 Wiring diagram EMGZ621A E621A0011e...
Page 26: ...Operating Manual EMGZ621A 26 10 Serial Interface RS232 Optional...
Page 31: ...Operating Manual EMGZ621A 31 13 Interface CAN Bus Option 14 Interface DeviceNet Option...