Operating Manual EMGZ621A
14
8.2
Configuring the Electronic Unit
Prior to the first calibration, the following settings must be done (ref. to „9. Parametrization“):
System parameters
Language
Required display language
Measuring system
Metric (default) or US standard
Parameters EMGZ621A
Force of sensor
Ref. to nameplate of the force sensor
Unit of sensor
Ref. to nameplate of the force sensor
Sensitivity
FMS force sensors = 1.8mV/V (default)
Scale output 1...4
Which tension feedback refers to 10V resp. 20mA?
Config. output
0-10V and 0...20mA
(default)
or
0-10V and 4...20mA
or
-10V...+10V
Notice: With a tension signal 0...10V, tension and current output
are active at the same time.
Jumpers for the analog outputs (ref. also to „14. Technical Reference“)
Channels 1 / 2 / 3 / 4
0...10V (Default) or ±10V,
corresponding to parameter
Config. output
Note
Wrong setting of the parameters may cause malfunction of the electronic unit! Setting of the
parameters must be done carefully prior to setting into operation!