Operating Manual EMGZ621A
20
9.4
Description of the system parameters
The parameter changing mode is activated by pressing the
PARA
↵
key for 3 seconds. By
pressing the
PARA
↵
key again, the system parameters are selected (ref. also to fig. 11).
Language
Use:
This parameter stores the display language.
Range:
English, French, Italian, German
Measuring system
Use:
This parameter indicates the measuring system to be used. If it is set to
metric
, all force values are shown as [N, cN, kN]. If it is set to
US
standard
, all force values are shown as [lb, clb, klb].
Range:
Metric, US standard
Default:
Metric
Lowpass display
Use:
The electronic unit provides a lowpass filter to prevent noise which is added
to the integrated display. This parameter stores the cut off frequency. The
lower the cut off frequency, the more sluggish the output signal will be. Due
to this filter, the value shown in the display will be much more stable in the
case of high fluctuations of the force value.
The lowpass display filter is independent to the other filters.
Range:
0.1
to
10.0
Default:
1.0
Increment:
0.1
Unit:
[Hz]
Identifier
Use:
This parameter stores the ident number of the device when linked to
PROFIBUS, CAN-Bus resp. DeviceNet.
Range:
0
to
255
Default:
0
Increment:
1
Unit:
[-]
Baud rate
Use:
This parameter stores the speed of the serial interface (RS232). The other
settings are fixed: 8 data bits, even parity, 1 stop bit („8 e 1“).
Range:
2400, 4800, 9600, 19200
Default:
9600
Unit:
[Baud]
Summary of Contents for EMGZ621A
Page 8: ...Operating Manual EMGZ621A 8 6 Dimensions fig 3 Dimensions E611002e...
Page 11: ...Operating Manual EMGZ621A 11 7 2 Wiring Diagram fig 6 Wiring diagram EMGZ621A E621A0011e...
Page 26: ...Operating Manual EMGZ621A 26 10 Serial Interface RS232 Optional...
Page 31: ...Operating Manual EMGZ621A 31 13 Interface CAN Bus Option 14 Interface DeviceNet Option...