Operating Manual EMGZ621A
21
9.5
Description of the Parameters EMGZ621A
The parameter changing mode is activated by pressing the
PARA
↵
key for 3 seconds. The
module
Params EMGZ621A 1
is then searched with the
↑
↓
keys and selected with the
PARA
↵
key (ref. also to fig. 11).
Offset A
Use:
This parameter stores the value determined with special function
Find
offset A
in [Digit]. It is not necessary to note this parameter because a new
offset adjustment is done very easy; also when changing the whole
electronic unit.
The offset can also be inputted manually with the
↑
↓
←
keys.
Range:
–8000
to
8000
Default:
0
Increment:
1
Unit:
[Digit]
Gain A
Use:
This parameter stores the value determined with special function
Calibration A
, or you must input a value calculated using the formulas
written under „8.3 Calibrating the measuring amplifier“, if the
material
tension
cannot be simulated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Gain 1 A
Use:
Identical
with
Gain A
but the value stored here was determined by special
fuction
Calib. gain 1 A
. The value stored here is used if the dig. input
„Switch gain A“ is activated (ref. to wiring diagram, fig. 6).
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Offset B
Use:
Identical
with
Offset A
but the value was determined by special fuction
Find offset B
.
Range:
–8000
to
8000
Default:
0
Increment:
1
Unit:
[Digit]
Gain B
Use:
Identical
with
Gain A
but the value was determined by special fuction
Calibration B
.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Summary of Contents for EMGZ621A
Page 8: ...Operating Manual EMGZ621A 8 6 Dimensions fig 3 Dimensions E611002e...
Page 11: ...Operating Manual EMGZ621A 11 7 2 Wiring Diagram fig 6 Wiring diagram EMGZ621A E621A0011e...
Page 26: ...Operating Manual EMGZ621A 26 10 Serial Interface RS232 Optional...
Page 31: ...Operating Manual EMGZ621A 31 13 Interface CAN Bus Option 14 Interface DeviceNet Option...