Installation
Mounting
27
5
Figure 5-1: Scan Line Width L and Distance to Target D
5.3 Mounting
The linescanner can be installed as follows:
•
on a tripod with a standard 1/4-20 UNC (photo equipment) thread. This type of setup requires the tripod
mounting plate (A-MP-MP, available as accessory) and is recommended if the linescanner is to be used
only for temporary or mobile measurements.
•
with the mounting holes of the linescanner housing. This type of setup is recommended for permanent
installations where higher stability is desired.
Prevent all contact between heat sources and the linescanner to protect the linescanner from overheating.
5.4 Warm-Up Time
For accurate temperature readings we recommend a 30 minute warm-up period after power on. During this time
the internal calibration sources will be stabilized. Digital communication and the analog outputs can be started
120 seconds after power on.