![Fluke MP Series Abridged User Manual Download Page 15](http://html.mh-extra.com/html/fluke/mp-series/mp-series_abridged-user-manual_2309932015.webp)
Technical Data
Measurement Specification
15
2
2 Technical Data
2.1 Measurement Specification
Temperature Range
MP150-3M
200 to 1500°C (392 to 2732°F)
MP300-3M
250 to 1500°C (482 to 2732°F)
MP150-2M
350 to 1500°C (662 to 2732°F)
MP300-2M
400 to 1500°C (752 to 2732°F)
MP150-1ML
600 to 1500°C (1112 to 2732°F)
MP300-1ML
650 to 1500°C (1202 to 2732°F)
MP150-1MH
700 to 1800°C (1292 to 3272°F)
MP300-1MH
700 to 1800°C (1292 to 3272°F)
Spectral Response
3M
2.4 µm
2M
1.6 µm
1ML
1 µm
1MH
1 µm
Detectors
3M
extended InGaAs
2M
InGaAs
1ML, 1MH
Si
System Accuracy
1
All models
± 0.5% of reading or ± 3°C (6°F), whichever is greater
Repeatability
2
All models
± 2°C (± 4°F)
Temperature Resolution
Digital output
0.1 K
Analog output
16 bit
Scan Rate
MP150
max. 150 Hz
MP300
max. 300 Hz
Measured Points per Line
MP150
256 pixel @ 150 Hz scan rate
512 pixel @ 80 Hz scan rate
1024 pixel @ 40 Hz scan rate
MP300
256 pixel @ 300 Hz scan rate
512 pixel @ 160 Hz scan rate
1
at ambient temperature 0 - 50°C (32 - 122°F), emissivity = 1.0 and calibration geometry
2
at ambient temperature 0 - 50°C (32 - 122°F), emissivity = 1.0 and calibration geometry