™/S, MOLBLOC TERMINAL OPERATION AND MAINTENANCE MANUAL
© 1995 - 2010 Fluke Corporation, DH Instruments Division
Page 10
Predicted Stability
2
(1 year)
±
0.05 % of reading
Measurement Uncertainty
3
With SP molbloc-S
calibration
A350K
±
0.2 % of reading from 50 to 200 kPa
A700K
±
0.2 % of reading from 50 to 500 kPa
S A1.4M
±
0.2 % of reading from 100 to 500 kPa
±
0.2 % of 100 kPa flow from 50 to 100 kPa
S A2M
±
0.2 % of reading from 200 to 500 kPa
±
0.2 % of 200 kPa flow from 50 to 200 kPa
Measurement Uncertainty
3
With LP molbloc-S calibration
A350K
±
0.2 % of reading from 20 to 200 kPa
A700K
±
0.2 % of reading from 50 to 200 kPa
±
0.2 % of 50 kPa flow from 20 to 50 kPa
Measurement Uncertainty
3
With HP molbloc-S
calibration
A700K
±
0.2 % of reading from 200 to 500 kPa
S A1.4M
±
0.2 % of reading from 200 kPa to 1.2 MPa
S A2M
±
0.2 % of reading from 200 kPa to 2 MPa
[1] Precision: Combined linearity, hysteresis, repeatability.
[2] Stability: Maximum change in zero and span over specified time period for typical and molbloc used
under typical conditions. As stability can only be predicted, stability for a specific molbloc and should
be established from experience.
[3] Measurement uncertainty (accuracy): Maximum deviation of the flow indication from the true value of
the flow through the molbloc including precision, stability and DHI calibration standard measurement
uncertainty.
1.2.5.2.2
Premium molbloc-S Calibrations
Measurement Update
Rate
1 second
Range
Depends on molbloc-S pressure dependent calibration type
(see Section 1.2.5.2.2)
Resolution
0.0015 % of FS
Linearity
±
0.05 % of reading
Repeatability
±
0.05 % of reading
Precision
1
±
0.06 % of reading
Predicted Stability
2
(1 year)
±
0.03 % of reading
Measurement
Uncertainty
3
With SP molbloc-S
calibration
A350K
±
0.125 % of reading from 50 to 200 kPa
A700K
±
0.125 % of reading from 50 to 500 kPa
S A1.4M
±
0.125 % of reading from 100 to 500 kPa
±
0.125 % of 100 kPa flow from 50 to 100 kPa
S A2M
±
0.125 % of reading from 200 to 500 kPa
±
0.125 % of 200 kPa flow from 50 to 200 kPa
Measurement
Uncertainty
3
With LP molbloc-S
calibration
A350K
±
0.125 % of reading from 20 to 200 kPa
A700K
±
0.125 % of reading from 50 to 200 kPa
±
0.125 % of 50 kPa flow from 20 to 50 kPa
Measurement
Uncertainty
3
With HP molbloc-S
calibration
A700K
±
0.125 % of reading from 200 to 500 kPa
S A1.4M
±
0.125 % of reading from 200 kPa to 1.2 MPa
S A2M
±
0.125 % of reading from 200 kPa to 2 MPa
[1] Precision: Combined linearity, hysteresis, repeatability.
[2] Stability: Maximum change in zero and span over specified time period for typical and molbloc used
under typical conditions. As stability can only be predicted, stability for a specific molbloc and should
be established from experience.
[3] Measurement uncertainty (accuracy): Maximum deviation of the flow indication from the true value of
the flow through the molbloc including precision, stability and DHI calibration standard measurement
uncertainty.