4. REMOTE OPERATION
Page 127
© 1995 - 2010 Fluke Corporation, DH Instruments Division
MFCCH
Purpose
Read or set the MFC channel and output mode (voltage or current).
Syntax
“MFCCH=channel,mode”
“MFCCH”
Defaults
“MFCCH=1,v”
Arguments
channel:
The optional MFC switchbox channel.
0
1
1 to 5
disables the MFC interface.
enables the optional MFC interface.
selects the MFC channel to use for optional MFC
switchbox
mode:
The MFC interface mode
‘v’
‘mA’
‘ ‘
Voltage control and measure mode.
The selected channel will be used for voltage
setting and measuring.
Current loop control and measure mode.
The selected channel will be used for current
loop setting and measuring.
If the argument is not given, the molbox will use
the selected channel to measure the voltage,
and will always use channel 1 to set the voltage.
This is used to measure voltage MFM’s that are
connected to channel 2 through 5 with an MFC
controlling on channel 1.
Remarks
The optional MFC interface can support voltage or current controlled MFC’s.
There is also an available MFC switchbox, which is controlled by the .
Example
Command:
Reply:
“MFCCH=1,
mA”
“1,mA”
Enables MFC for current
Command:
Reply:
“MFCCH=3,
V”
“3,V”
Enables MFC on switchbox channel 3 for volts
Command:
Reply:
“MFCCH=0”
“0,V”
Disables MFC interface
Command:
Reply:
Command:
Reply:
“MFCCH=1,
V”
“1,V”
“MFCCH=4”
“4,V”
Enables MFC control on channel 1
Enables MFM input on channel 4
Errors
ERR# 6:
ERR# 23:
ERR# 23:
Channel or mode arguments are invalid
The MFC option is not installed
Channel greater than 1 and MFC switchbox option is not installed
See Also
3.4.8, 7.3