8.2.2
Calibration Commands
The following group of commands directly or indirectly affect the accuracy of
the instrument. These commands should only be used by qualified, authorized
personnel.
Calibration parameters can only be changed if protection is dis-
abled by entering the proper pass-code using the command *pas= ddddd.
Entering an invalid passcode or cycling the power re-establishes protection of
the calibration parameters. The parameters can be read without disabling
passcode protection.
Description
Command
Example
Response
Format
Response
Example
Range
Unprotect calibration parame-
ters with the pass-code
*pas[scode]=nnnnn
*pas=12345
Read passcode protection
state
*pas
*pas
ps: 9
Ps: 0
0 or 1
Read date
da[te]
date
da: yyyy-mm-dd
da: 1999-03-28
Set date
da[te]=yyyy-mm-dd
da=1999-03-28
1999-01-01 to
2100-12-31
Read probe serial number
se[rial]
ser
ser: xxxxxxxxxx
ser: 100231A
Set probe serial number
se[rial]=xxxxxxxxxx
ser=001
0 to 9, A to Z,
., -, /
Read probe type
p[robe]
pr
pr: {ITS-90, IEC751,
CVD, YSI400, THERM}
pr: ITS-90
Set probe type
p[robe]=IT[S-90]/
IE[C751]/ C[VD]/
Y[SI400]/ T[HR]
pr
ITS-90,
IEC751,
CVD, YSI400,
THERM
Read ITS-90 R(.01) or CVD R0 r[0]
r0
r0: 999.999
r0: 100.035
Set ITS-90 R(.01) or CVD R0
r[0]=n
r0=100.106
1 to 900
Read ITS-90 coefficient
a/b/c/d/a4/b4
a
{a/b/c/d/a4/b4}:
-9.999999E-99
a: -2.8644101E-05
Set ITS-90 coefficient
a/b/c/d/a4/b4=n
a=-2.8644E-5
1.0 to 9.9e-18
Read CVD coefficient
al[pha]/de[lta]/be[ta]
al
{al/de/be}: 9.99999999 al: 0.00385762
Set CVD coefficient
al[pha]/de[lta]/be[ta]=n al=0.0038512
Coefficient
Dependent
Read THERM thermistor
coefficient
b0/b1/b2/b3
b3
{b0/b1/b2/b3}:
9.999999E-99
b3: -9.944070E-12
Set THERM thermistor
coefficient
b0/b1/b2/b3=n
b3=-9.944E-12
E+/-18