At least one channel can be activated for data logging per transmitter. Therefore, at least
one channel is mandated for selection before a
Task
can be considered valid.
•
Internal (Integral) Channel
If applicable, the
Internal (Integral) Channel
can be selected. This channel is either integral
or, in the case of Thermocouple variants, is available once a Lumberg probe is connected.
•
External Channels (1-4)
If applicable,
External Channels (1-4)
can be selected. This is transmitter model dependent
as some transmitter models channels 1 and 2 may be integral to the transmitter.
•
Door Sensor
If applicable, the
Door Sensor
channel can be selected.
All available channels are eligible for
Alarm Programming
. Alarm options are as follows:
•
Temperature and Temperature/Humidity Channels
The available alarms for each channel include
Low Alert
and
Low Alarm
,
High Alert
and
High Alarm
. Alarms are optional and are all independent of each other.
NOTE
- If selected, values will be cross-checked with each other as well as with the limits
associated with the channel selected. This ensures all programmed values are valid.
•
Alarm Delay (Optional)
Alarm Delay is simply a time that will delay the action of raising the alarm and helps
overcome unnecessary alarm triggers for measured values that hover around the alarm
value.
Scale Mapping
is unique to the WiFi Multi-Parameter Transmitter (RF615). Each external
channel can use a preferred
Scale
(i.e., Pascals or Lumens). The
Scale Mapping
must be
correctly entered for the appropriate transducer connected.