9118A
Users Manual
64
Calibration Procedure for Tube
Furnace
Caution
The Product can be used with or without an
isothermal block installed. Unique calibration
parameters are necessary for each mode of
operation. To prevent Product damage, before
using the Product, select the correct mode of
operation in the calibration menu (see Furnace
Setup Menu). If necessary, a calibration for the
new mode of operation may be required.
1. Connect the test equipment as shown in Figure 25.
Note
In the steps that follow, start the calibration
procedure at the lowest temperature (300
°
C) and
move up the range to 1200
°
C.
2. Set the Product to one of the setpoint temperatures in
Table 8.