Calibration Furnace
Alphabetic List of Commands
47
SOURce:LCON:PBAN <n>
Sets the main control proportional band in
°
C which the PID controller uses for main zone control.
Parameters: n =
1.0
to
199.9
°
C
Example:
SOUR:LCON:PBAN 7
SOURce:LCON:PBAN?
Gets the main control proportional band in
°
C which the PID controller uses for main zone control.
Example:
SOUR:LCON:PBAN?
Response:
1.5
The main control loop proportional band is 1.5
°
C.
SOURce:LIST:SPO<i> <n>
Sets the temperature value for a preset setpoint in the RAMP/SOAK menu.
Parameters:
i
= Setpoint number (1 to 8)
n
=
Setpoint
temperature.
Example:
SOUR:LIST:SPO6 700.00
Sets setpoint number 6 to 700.00 degrees.
SOURce:LIST:SPO<i>?
Gets the temperature value for a preset setpoint in the RAMP/SOAK menu.
Example:
SOUR:LIST:SPO4?
Response:
400.0
The setpoint temperature for setpoint 4 is 400.0 degrees.
SOURce:PROT:HCUT?
Gets the hard cutout temperature setpoint in
°
C or
°
F.
Example:
SOUR:PROT:HCUT?
Response:
140
The hard cutout temperature is set to 140
°
C.