9118A
Users Manual
46
SOURce:LCON:DER <n>
Sets the main control derivative time in seconds.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameters: n
0.0
to
99.9
seconds.
Example:
SOUR:LCON:DER 5
SOURce:LCON:DER?
Gets the main control derivative time in seconds.
Example:
SOUR:LCON:DER?
Response:
1.5
The main zone derivative time is 1.5 seconds.
SOURce:LCON:INT <n>
Sets the main control loop integral time in seconds.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameter: n =
10.0
to
999.9
seconds.
Example:
SOUR:LCON:INIT 10
SOURce:LCON:INT?
Gets the main control loop integral time in seconds.
Example:
SOUR:LCON:INIT?
Response:
20.0
The main control loop integral time is 20.0 seconds.