Calibration Furnace
Alphabetic List of Commands
51
n
=
-10.00
to
10.00
Example:
SOUR:SENS:CAL:SKEW2 0.08
SOURce:SENS:CAL:SKEW<x>?
Gets an axial compensation adjustment parameter.
Example:
SOUR:SENS:CAL:SKEW2?
Response:
0.05
The axial compensation for SKEW2 is 0.05 of the main heater power.
SYST:CONF:BLC [n]
Sets the furnace work mode
with the isothermal block inside or without the isothermal block inside
.
Parameters: n =
0
determines if the furnace will work without the isothermal block inside
=
1
determines if the furnace will work with the isothermal block inside
Example: SYST:CONF:BLC 1 Determines if the furnace is to work with the isothermal block inside.
SYST:CONF:BLC?
Gets the furnace work mode
with the isothermal block inside or without the isothermal block inside
.
Example: SYST:CONF:BLC?
Response: 0 The furnace works without the isothermal block inside
SOURce:SENS:DATA? [heater part]
Gets the control temperature of the parts of the furnace.
Parameters: heater
part
MAIN
– gets main heater temperature (Default)
ZONE
– gets zone heater temperature
FULL
– gets main and zone heater temperature
Example:
SOUR:SENS:DATA? MAIN