7526A
Calibration Manual
20
Use the 742A-1K for the 400
Ω
range. Use the 742A-10K for the 4 K
Ω
range. For more
information about how to use the ratio mode, see the
8508A Operators Manual
. Note the
ratio indication on the 8508A in Table 8. Use the formula: (742A actual value/ratio
indication x 100 = actual 5522A actual value). Start with the 7526A in the 400
Ω
range.
After the 400
Ω
test, change the 7526A to the 4000
Ω
range.
Table 8. Ohms Ratio Table
Range
742A
Value
(
Ω
)
Ratio
5522A
Actual
Value (
Ω
)
7526A
Reading
90-Day
Specification
1-Year
Specification
0
Ω
0.0035
Ω
0.004
Ω
25
Ω
0.004
Ω
0.0045
Ω
75
Ω
0.005
Ω
0.0055
Ω
95
Ω
0.0054
Ω
0.0059
Ω
100
Ω
0.0055
Ω
0.0060
Ω
105
Ω
0.0056
Ω
0.0061
Ω
150
Ω
0.0065
Ω
0.007
Ω
300
Ω
0.0095
Ω
0.010
Ω
400
Ω
0.0115
Ω
0.012
Ω
0
Ω
0.035
Ω
0.04
Ω
250
Ω
0.04
Ω
0.045
Ω
750
Ω
0.05
Ω
0.055
Ω
1000
Ω
0.055
Ω
0.06
Ω
1500
Ω
0.065
Ω
0.07
Ω
2000
Ω
0.075
Ω
0.08
Ω
3000
Ω
0.095
Ω
0.10
Ω
4000
Ω
0.115
Ω
0.12
Ω