Precision Process Calibrator
Performance Tests
19
Ohms Input
The ohms input performance test verifies the ohms input on the front of the Product.
Before an ohms input is measured, the ohms output of a 5522A must be “characterized”.
To get the necessary accuracy, the 8508A is used as a transfer standard and the 742A is
used as the reference standard. To find the true value of the 5522A output, use the ratio
input function of the 8508A. This is located on the front and rear panels. Connect the
742A to the front terminal and connect the 5522A to the rear terminals. Use four-wire
measure for both. See Figure 7 for connections.
TRIG
5
V
PK MAX
REAR I
N
PUT
CURRE
N
T
FUSE.
F 1.6AH
5x20mm
IEC127
I
N
PUT (2
W
IRE)
HI
LO
GU
3A PK MAX
920
V
PK MAX
1420
V
PK MAX
LO
HI
CALIBRA TIO
N
SE
N
SE (4
W
IRE )
E
N
ABLE
DISABLE
A
CAT I @ 1000
V
742A
8508A Front
8508A Rear
Sense Hi
Hi
Lo
Sense Lo
5520A
Sense Hi
Hi
Lo
Sense Lo
gxm015.eps
Figure 7. Ohms Input Performance Test Connections
Summary of Contents for 7526A