DC Current
The dc current performance test verifies the accuracy of the current output from the front
panel of the Product. Use the 8508A and the precision shunt to measure the Product
output. Divide the voltage indication from the 8508A by the 742A-1 actual value to
determine the current output. Verify the measurements shown in Table 5. See Figure 2
for connections.
Table 5. DC Current Verification
100 mA
Output
Volt
Current
(I=E/R)
742A-1 Shunt
Value
90-Day
Specification
1-Year
Specification
0.000 mA
Direct into
8508A
Current Input,
Autorange
1.0
μ
A
1.0
μ
A
35.000 mA
2.4
μ
A
2.75
μ
A
65.000 mA
3.6
μ
A
4.25
μ
A
100.000 mA
5
μ
A
6
μ
A
CURRENT
OUTPUT
HI
VOLTS
mA
RTD/
OUTPUT
HI
LO
HI
LO
LO
SENSE
4W RTD/
INPUT
20 V PK
MAX
100 V MAX
100 mA MAX
20 V PK
MAX
PRECISION PROCESS CALIBRATO
7526A
8508A
Current Shunt
7526A
gxm010.eps
Figure 2. DC Current Performance Test Connections