7526A
Calibration Manual
16
CJC (Cold Junction Compensation)
The CJC performance test verifies the accuracy of the cold junction compensation of the
Product. Connect a Type-E thermocouple to the TC terminals on the Product. Immerse
the thermocouple and a precision thermometer in a mineral oil lag bath. See Figure 4 for
connections.
1.
Push
then
to switch to TC input.
2.
Push
then
to turn on the internal junction. XCJC is not shown when the
internal junction is used.
3.
Push
until Type E is shown on the display.
Make sure that the indications of the Product and the precision thermometer are within
the Type-E specifications:
•
90-Day specification is 0.06
°
C
•
1-Year specification is 0.08
°
C
CURRENT
OUTPUT
HI
VOLTS
mA
RTD/
OUTPUT
HI
LO
HI
LO
LO
TC
SENSE
INPUT/OUTPUT
4W RTD/
INPUT
20 V PK
MAX
100 V MAX
100 mA MAX
20 V PK
MAX
20 V PK
MAX
PRECISION PROCESS CALIBRATOR
7526A
Mineral Oil
Lag Bath
Dewar Flask
and Cap
HI
Precision
Thermometer
7526A
gxm012.eps
Figure 4. CJC Performance Test Connections
Note
Typical Type-E thermocouples do not have specifications accurate enough
to be used as a standard. To maintain a good test uncertainty ratio (TUR),
a characterized Type-J thermocouple can be used.