8588A/8558A
Remote Programmer's Manual
48
The Transition register is not used in the Product. The state of the other three
registers is accessed with the STATus commands as shown in Table 10.
Table 10. STATus Commands
Command
Description
STATus
:OPERation
:CONDition?
Returns the contents of the Operation
Condition register. The only
OPERational:CONDition sub-register
implemented in the Product is that of Bit 4
“MEASuring”. This sub register is false if the
Trigger system is in the Idle state; true if the
trigger system is initiated.
[:EVENt]?
Set true if the event has occurred. Only used in
bits 4, 5, 6, 8, 9, 10, and 11.
:ENABle <nrf
>
Set which OPERational register bits cause the
OPER bit (7) in the Status byte to be set true.
Nrf
is the sum of bit values to be set; for
example:
nrf
= 80 set bits 4 and 6 to enabled.
Min= 0, Max = 32768
:ENABle?
Returns the sum of the bit values of the
enabled bits.
:PRESet
Sets Enable bits both OPERational and
QUEStionable registers to False.
:QUEStionable
:CONDition?
The Condition sub register is not used in the
Product
[:EVENt]?
Set true if the event has occurred. Not used in
bits 0, 1, 2, 6, 7, 8, 13, 14 and 15
:ENABle <
nrf>
Set which QUEStionable register bits cause the
QUES bit (3) in the Status byte to be set true.
Nrf
is the sum of bit values to be set; for
example:
nrf
= 2056 set bits 3 (8) and 11 (2048)
to enabled. Min = 0, Max = 32768
:ENABle?
Returns the sum of the bit values of the
enabled bits