Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
35
Table 6. Command Summary (cont.)
Keyword
Description
:SECondary?
Returns the selected secondary reading setting
SECondary:METHod
MEASure|SINE|SQUare|TRIangle|TSINe
Set the method by which the secondary peak-to-peak value is
calculated
MEASured = The measured positive peak minus the measured
negative peak value
SINe = The peak to peak of a sine wave calculated from the rms
value
SQUare = The peak to peak of a square wave calculated from
the rms value
TRIange = The peak to peak of a triangle waveform calculated
from the rms value
TRUNcated = The peak to peak of a truncated sine wave
calculated from the rms value
:SECondary:METHod?
Returns the method by which the secondary peak-to-peak value
is calculated
[:SENSe]:VOLTage:DC
:APERture
<nrf>
|MIN|MAX|DEF
Sets the ADC aperture value in seconds or to the MIN, MAX, or
Default setting
:APERture? [MIN|MAX|DEF]
Returns the aperture setting or the MIN, MAX, or Default
settings in seconds
:APERture:MODE AUTO|FAST|MANual
Sets the aperture mode
:APERture:MODE?
Returns the aperture mode
:NPLC <plcs>MIN|MAX|DEF
Sets the ADC aperture in number of power line cycles or to the
MIN, MAX, or Default plc setting
:NPLC? [MIN|MAX|DEF]
Returns the ADC aperture in number of power line cycles or to
the MIN, MAX, or Default plc setting
:IMPedance AUTO|1M|10M
Sets the input impedance. Auto is > 1 T
Ω
if range is < 100 V,
otherwise 10 MΩ
:IMPedance?
Returns the input impedance setting
:RANGe:AUTO[:STATe]
<boolean>
Turns Auto range ON or OFF
:RANGe:AUTO[:STATe]?
Returns 1 for Auto range ON, 0 for auto range OFF
:RANGe
<nrf>
|MIN|MAX|DEF
Set maximum expected value or min, max or default range