Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
19
Commands by Subsystem
Table 6. Command Summary
Keyword
Description
CALCulate
CALCulate:AVERage
AVER controls values displayed and returned by FETCH?
:STATe
<boolean>
Turns State ON or OFF. Default OFF.
:STATe?
Return the state of AVERage
:COUNt<
nrf
>
Set the number in an AVERage. Default 16, Min 1, Max 10,000.
:COUNt?
Returns the number of readings averaged
:TCONtrol MOVing|REPeat
MOVing = rolling; REPeat = Block. Default MOVing.
:TCONtrol?
Returns the control setting (MOV or REP)
CALCulate:LIMit
:CLEar:[IMMediate]
immediately clear all flags
:LOWer
<nrf>
Set the lower limit. Default -1.9915
:LOWer?
Return the ower limit value
:LOWer:STATe
<boolean>
Turn the lower limit ON or OFF. Default OFF.
:LOWer:STATe?
Return the lower limit state
:FAIL?
Returns: Outside limits = 1, within limits = 0
[:STATe]
<boolean>
Turn Limits ON or OFF. Default OFF.
[:STATe]?
Return the limits state
:UPPer
<nrf>
Set the upper limit. D1.999999999e
:UPPer?
Return the lower limit value
:UPPer:STATe
<boolean>
Turn the upper limit ON or OFF. OFF.
:UPPer:STATe?
Return the upper limit state
CALCulate:MATH
Math calculates (mx - c)/z
:CFACtor
<nrf>
|LREading
Sets the c variable to a value or the last reading. Default 1, Min -
9.9E+37, Max 9.9E+37.
:CFACtor?
Returns the c variable value
:CFACtor:STATe
<boolean>
Turns the - c operation ON or OFF. Default OFF.
CFACtor:STATe?
Returns the state of c operation (I = On, 0 = Off)
:MFACtor
<nrf>
|LREading
Sets the m variable to a value or the last reading. Default 1, min
-9.9E+37, Max 9.9E+37.
:MFACtor?
Returns the m variable value
:MFACtor:STATe
<boolean>
Turns the multiply by m operation ON or OFF. Default OFF.
:MFACtor:STATe?
Returns the state of m operation (I = On, 0 = Off)