8588A/8558A
Remote Programmer's Manual
54
Keyword
Description
ARM[:SEQuence]:LAYer2
ARM[:SEQuence]:LAYer1
(The two ARM layers have identical commands)
:COUNt
<nrf>
Specifies the number of passes through
layer
and subordinate
layers before control flows up to the superior layer. Default = 1,
Max = 10000000
:COUNt?
Returns the Count setting
:COUPling DC|AC
Sets the coupling used with :SOURce:INTernal. See also :LEVel
and :SLOPe. Default = AC
:COUPling?
Returns the coupling setting
:DELay:AUTO
<boolean>
When ON, the time delay between
source
event
detection and
flow passing to the subordinate layer is determined
automatically (the DMM sets an appropriate delay needed for
settling after a configuration change). Default = ON
:DELAY:AUTO?
Returns the auto delay setting, 1 = ON, 0 = OFF
:DELay
<nrf>
Set the time delay between
source
event
detection and flow
passing to the subordinate layer.
:DELay?
Returns the delay setting in seconds
:ECOunt
<nrf>
numeric
is an integer value to specify how many ARM events
must be recognised before DELay is started on the downwards
transit through the layer. ECOunt provides ARM event division.
Default = 1, Max = 10000000
:ECOunt?
Returns the Ecount setting
:EXTernal
POSitive|NEGative
[,TTL|BIPolar]
Sets the polarity of the External trigger edge and optionally
signal type to TTL or Bipolar. Default = Negative, TTL
:EXTernal?
Returns the external edge polarity and edge type settings
:FILTer
<boolean>
Turns the filter in the event detector path ON or OFF. The filter
affects both the signal and event detector paths. Default = OFF
:FILTer?
Returns the filter state
:IMMediate
Modifies event detection behaviour. If the system is waiting for
an event in this layer, this one-time command bypasses event
detection ECOUNT and DELay causing immediate exit from the
layer on the downward path. If not waiting at the event detector
the command is ignored and error -221 issued.
:LEVel
<nrf>
Sets the percentage of range at which arming occurs when
Source is INTernal. Default = 0%, Min -200 %, Max = 200 %